Inderm Nappage: Features

 

Device Components :

 


__________________________________________________________________________

Explanation of Buttons:

__________________________________________________________________________

Manual Operation Mode:

Select Mode:
  • Continuous
  • Micro
  • Standard
  • Auxiliary
    Cycle: (syringe assembly) moves forward to release the contents of the syringe
  • Nappage

Continuous Mode: The gun will cycle while the trigger is held, the cycle will stop when the trigger is released.

Micro Mode: The gun will cycle one time when the trigger is activated and released. The Micro Mode is used for short duration cycling. The duration of the cycle is adjustable by using the knob on the rear panel, clockwise will increase the cycle time; counter clockwise will decrease the cycle time.

Standard Mode: The gun will cycle one time when the trigger is activated and released. The duration of the cycle is adjustable by using the knob on the rear panel, clockwise will increase the cycle time; counter clockwise will decrease the cycle time.

Auxiliary Mode: Available for customer customization.

Nappage Mode: The gun will cycle on and off as long as the trigger is held, releasing the trigger will end the cycle. The duration of the cycle is adjustable by using the knob on the rear panel; clockwise will increase the time between cycle, counter clockwise will decrease the time between cycle time.

Pre-Sets: There are four pre-sets that can be stored in the gun; P1, P2, P3, P4. Settings can be saved or changed at any location you wish. Note that if all power is removed from the gun the pre-sets are lost and must be reentered.

__________________________________________________________________________

Setting Pre-Sets: In the manual mode find a setting you wish to make a pre-set. Once you have found the settings that best suit what you want, press the set button and hold until the location where you want to save the pre-set is selected, release the button and the led will remain lit, press the trigger the led will flash the gun will beep the led goes off, release the trigger, the pre-set is saved at that location. Using the same procedure you can overwrite an existing pre-set.

NOTE: If you decide to not save a new pre-set, not to overwrite a current pre-set or want to exit pre-set mode altogether, press and hold the manual button and release on your mode of choice or turn off the device.

  • Set Pre-Sets from the Micro Mode, Standard Mode or the Auxiliary Mode only.
  • Use only B&D Luer Lock® syringes only; 1cc, 3cc, 5cc or 10cc.
  • Make sure that the syringe plunger lock is in the closed position.
  • Use only 4mm and 6mm needles. Not advised for use with needles over 6mm.
  • Always set the needle depth guard before using.

Note: Gun may only be used in a horizontal or downward position.

 
 