Inderm PSI: Instructions

 
Prior to using the Inderm-PSI carefully read all instructions. Remove the Inderm from the canvas case and place on a flat surface.

 
Changing the Battery

To charge the Inderm, remove the bottom cap of the Inderm device by rotating it 45 degrees counter clockwise. Unplug the battery from the battery receptacle located at the bottom of the device as shown to the left. Remove the battery by pulling on the 3 prong adaptor, do not pull battery by the wires of the battery itself. Replace the old battery with a fully charged battery. Make sure the 3 prong adaptor snaps into the receptacle on the bottom of the device securely.
 
Charging the Battery

To charge the battery, place the battery in the cradle of the battery charger. Take the 3 prong adaptor (on the battery) and plug it into the battery charging receptacle (shown below with the red arrow) . Then, plug the wall mount adaptor into the backside of the battery charger and plug the other side into a wall outlet. The light on the battery charger will glow red to show that the battery is charging. Over time the light will change to yellow when the battery is at half charge.
  Determining when the Battery is Fully Charged
When the battery charger light is blinking yellow / green the battery is almost fully charged. When the light turns to solid green the battery is fully charged. While the light glows solid green, the battery may be removed from the battery charger and then plugged back into the Inderm device.
 
Syringe Cradle

The syringe cradle is provided in two different offset settings. The Z-shaped cradle is needed for full 10cc syringes. The 10cc syringe will only accommodate an 8cc fill when used with the straight cradle. The straight cradle is used with 1, 3, and 5cc syringes. Depending on the length of the syringe needle, the syringe cradle should be exchanged so that the needle does not protrude beyond the needle guard.
In addition, the depth that the needle will penetrate the skin can be adjusted using the depth-adjusting knob. This adjustment affects how far the needle guard will travel into the housing once activation occurs and thus how deep the needle will penetrate the skin, it will not result in a shift of the needle guard until activation occurs. This depth adjustment can be viewed via a shift of the red indicator located in a transparent window.

 
To install the syringe cradle, first unscrew the mounting screw from the syringe cradle. Then place the syringe cradle on the cradle mounting bracket and tighten the mounting screw.

Above the wheel, on the top of the device, is a clear window with a red indicator inside, this indicates the approximate depth of the needle guard as there will not be a visible shift of the needle guard. The needle guard has a depth adjustability of over a half an inch.
   